Health Products

ALB2 (Albumin Gen. 2) for Cobas C701 Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Troche Diagnostics Ltd.
ALB2 (Albumin Gen. 2) for Cobas C701 Analyzer05166861190

Introduction

Recalled Products:

ALB2 (Albumin Gen. 2) for Cobas C701 Analyzer

Reason

Is related to a possible issue where patient and quality control albumin results may become erratic over time with a continuous throughput of more than 300 tests in a shift with deviations of up to 10%.