Eclipse Treatment Planning System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Varian Medical Systems Inc.
Eclipse Treatment Planning System100042024

Introduction

Recalled Products:

Eclipse Treatment Planning System

Reason

Eclipse treatment planning system where an outdated assigned Hounsfield Unit [Hu] value can be used for dose calculation when the clock on the Eclipse Client workstation is not synchronized with the database / system server. When the Eclipse client workstation clock is running behind the clock on the database / system server, it is possible that a change to the assigned HU value for a structure does not invalidate the density image stored in the cache, and for the subsequent dose and Monitor Unit [MU] calculation to be based on the outdated HU assignment. The resultant MUs may be higher or lower than expected, dependent on the specific HU assignments. Treatment of the patient using these values can therefore lead to under or over dose.