Health Products

Haloflex Energy Generator

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices, Product Safety
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
010410-001, 010410-002, 090909-009, 100711-003, 100711-004, 100711-005, 111209-009, FE103845.20
Manufacturer
Barrx Medical Inc.
1. Haloflex Energy Generator1190A-115A

Introduction

Recalled Products

1. Haloflex Energy Generator

Reason

There is a generator malfunction where Haloflex Energy Generator failed to enter "Standby mode" when initially powered-on (plugged in and turned on).  During normal use when the device is powered-on, the generator automatically sequences through a "Power-on self test" to check for proper function.  If the generator passes the self- test, it enters standby mode, displays "Ready, connect catheter" and is ready for use.  Patient injury has not been seen with this malfunction.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime