Health Products

ImmunoCard Mycoplasma

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
709030C714, 709030C713
Manufacturer
Meridian Bioscience Inc.
1. ImmunoCard Mycoplasma709030

Introduction

Recalled Products

1. ImmunoCard Mycoplasma

Reason

Meridian has received reports from customers of invalid negative control reactions and of blue-coloured substrate reagent. As indicated in the product package insert, substrate reagent should be colourless. Substrate reagent exhibiting a blue colour should be discarded. Preliminary investigation indicates that the invlaid negative control reactions are attributable to the blue substrate reagent (lot 7801.486). The remaining kit components and reagents are meeting performance specifications.