Health Products

Elecsys FT4

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices, Product Safety
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
16519303, 16519304, 16649801, 16649803
Manufacturer
Roche Diagnostics GMBH
A. Elecsys FT411731297122

Introduction

Recalled Products
  1. Elecsys FT4

Reason

During recent internal on-board stability studies, decreasing calibrator signals were observed for Elecsys FT4 reagent kits when stored more than 3 weeks on-board the Elecsys systems (Cobas E 411 and Cobas E 601/602)