AU5800 Clinical Chemistry Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All Lots
Manufacturer
Beckman Coulter Inc.
A. AU5800 Clinical Chemistry AnalyzerAU5810

Introduction

Recalled Products
  1. AU5800 Clinical Chemistry Analyzer

Reason

If the AU5800 diluted detergent tank valve is positioned to "closed" during the instrument "measure" mode, diluted detergent will not be dispensed and may adversely affect washing of cuvettes, probes and mixing-bars.