D. CRPL3 (C-Reactive Protein Gen.3) for Cobas C701 Analyzer
Manufacturer
Roche Diagnostics Ltd.
05172373190
All Lots
E. Ureal (Urea/Bun) for Cobas C701 Analyzer
Manufacturer
Roche Diagnostics Ltd.
05171873190
All Lots
B. CO2-L (Bicarbonate Liquid) for Cobas C701 Analyzer
Manufacturer
Roche Diagnostics Ltd.
05446376190
All Lots
C. CREP2 (Creatinine Plus Ver.2) for Cobas C701 Analyzer
Manufacturer
Roche Diagnostics Ltd.
05168597190
All Lots
A. ALP2 (Alkaline Phosphatase AC. To IFCC Gen.2)
Manufacturer
Roche Diagnostics Ltd.
05166888190
Introduction
Recalled Products
ALP2 (Alkaline Phosphatase AC. To IFCC Gen.2)
CO2-L (Bicarbonate Liquid) for Cobas C701 Analyzer
CREP2 (Creatinine Plus Ver.2) for Cobas C701 Analyzer
CRPL3 (C-Reactive Protein Gen.3) for Cobas C701 Analyzer
Ureal (Urea/Bun) for Cobas C701 Analyzer
ALT (Alanine Aminotransferase) IFCC"
Reason
Due to a software issue on the Cobas 8000 Analyzer, a mismatch between the reagent probe positioning on the Cobas C701 and C702 Analyzers and the aspirated reagent volumes may occur under certain high throughput conditions. This may lead to partial air pipetting at the end of the Cobas C Pack large, possibly affecting patient sample and control recovery.