Health Products

Access system recalls

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 Numbers, Contact manufacturer
Manufacturer
Beckman Coulter Inc.
F) Synchron LXI 725 Clinical System476501, A15642
More than 10 Numbers, Contact manufacturer
Manufacturer
Beckman Coulter Inc.
D. Access 2 Immunoassay System386220, 81600N
More than 10 Numbers, Contact manufacturer
Manufacturer
Beckman Coulter Inc.
E) Access 2 Immunoassay System  Part Numbers 81600N, 386220386220, 81600N
More than 10 Numbers, Contact manufacturer
Manufacturer
Beckman Coulter Inc.
B. Unicel DXC 600I Synchron Access Clinical System4767, A25640
More than 10 Numbers, Contact manufacturer
Manufacturer
Beckman Coulter Inc.
C. Access Immunoassay System Part Number 8160081600
More than 10 Numbers, Contact manufacturer
Manufacturer
Beckman Coulter Inc.
A. Access Immunoassay System81600

Introduction

Recalled Products
  1. Access Immunoassay System
  2. Unicel DXC 600I Synchron Access Clinical System
  3. Access Immunoassay System Part Number 81600
  4. Access 2 Immunoassay System 
  5. Access 2 Immunoassay System  Part Numbers 81600N, 386220
  6. Synchron LXI 725 Clinical System

Reason

Potential problem with the 10-Tooth (10T) and 14-Tooth (14T) Mixer Pulley Assemblies in the access family of Immunoassay systems. It has been determined that in some cases, pulley drive components have broken parts or have become loose, mis-positioned, or disconnected from their drive. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate mixing of the contents of the reaction vessel which can lead to erroneous test results.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime