Subject: ZOLL AED Plus Defibrillator May Not Deliver Therapy
Dear Health Care Professional,
We have received reports that following a long period (typically greater than four years) without use, a ZOLL® may prompt "change batteries" during use and fail to deliver therapy. Turning the device off completely, waiting ten seconds for the unit to re-set and then turning it back on is necessary to resume proper operation. Our investigation of this problem has determined that some batteries may develop high internal resistance that interferes with the batteries expected performance, lengthening charging time beyond specified and clinically acceptable limits.
ZOLL Medical Corporation is informing owners of all AED Plus devices with serial numbers below X_ _ _200000 to replace batteries that have been installed in devices for a period longer than three (3) years and at subsequent three (3) year intervals.
Units above serial number X_ _ _200000 are not affected by this corrective action as they contain software that can detect this battery condition and identify when batteries require replacement. In lieu of adding the battery replacement label users may obtain this new software for their AEDs at no cost at ZOLL AED Plus Battery Help.
ZOLL AED Plus Battery Help will be available on February 16th 2009 and will provide other technical information and support concerning this corrective action, such as help in determining battery age and ordering of replacement batteries and extra labels. We have arranged special pricing of $39.95 per set of ten batteries inclusive of shipping that can be ordered by credit card at this site. We are offering this special pricing for customers affected today as well as for future replacements to mitigate the cost associated with a shorter than expected battery life.
Health Canada has been notified of this corrective action and classified it as a Priority II recall.
We apologize for any inconvenience this may cause you and thank you in advance for assistance in implementing this corrective action. Avoiding this problem during clinical use is our highest priority. Our 24/7 technical support numbers 1 (800) 348-9011 or+1 (978) 421-9460 are available to assist users with any aspect of this notice.
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-marketing adverse incidents are generally presumed to underestimate the risks associated with health product treatments. Any cases of serious specific adverse incident or other serious or unexpected adverse incidents in patients using ZOLL AED Plus Defibrillator should be reported to ZOLL Medical or Health Canada at the following addresses:
ZOLL Medical Corporation
Worldwide Headquarters
269 Mill Road
Chelmsford, Massachusetts 01824-4105
Tel.: 1 (800) 348-9011 or +1 (978) 421-9460
Any suspected adverse reaction can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada web site.
For other inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate (MHPD)
E-mail: MHPD_DSPC@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
Sincerely,
original signed by
Paul Dias
VP Quality Assurance & Regulatory Affairs
Addendum to the ZOLL AED Plus Administrator's Guide
Applies to all units with SW version 5.18 and lower
Please Note: Replace the batteries in your AED Plus® three years from the installation date of new batteries.
Explanation of Symbols:
The Replace batteries before date should go at the right of the battery and hourglass symbol - Replace batteries before date
Place the AED Battery Replacement Label below the AED Plus Status Indicator
AED Battery Replacement Label located below the AED Plus Status Indicator
9652-0120-01, Rev A