B. Beckman Coulter AU 5400 Immunochemistry System - Urinary/CSF Protein
OSR6170, OSR6270
2796, 2812
Manufacturer
Beckman Coulter Inc.
C. Beckman Coulter AU Immunochemistry System - Urinary/CSF Protein
OSR6170, OSR6270
3241, 3257
Manufacturer
Beckman Coulter Inc.
A. Beckman Coulter AU2700 Immunochemistry System - Urinary/CSF Protein
OSR6170, OSR6270
Introduction
Recalled Products
Beckman Coulter AU2700 Immunochemistry System - Urinary/CSF Protein
Beckman Coulter AU 5400 Immunochemistry System - Urinary/CSF Protein
Beckman Coulter AU Immunochemistry System - Urinary/CSF Protein
Reason
Beckman Coulter has confirmed that a negative shift in quality control and patient sample recovery may be observed at the low end of the measuring range, resulting in an increase in negative results with G flags, when certain device lot numbers are used. Additionally, a blue precipitate has been observed in some reagent vials of these lots.