DRB1*04 SSP UniTray Kit

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
010 666807
Manufacturer
Life Technologies Corp. (Invitrogen)
A. DRB1*04 SSP UniTray Kit45040-4

Introduction

Recalled Products

A. DRB1*04 SSP UniTray Kit

Reason

One complaint regarding this issue has been received. Complaint states the DRB1*04 SSP Unitray Kit has a false negative in lane 9 which gives no perfect match typing result for the DRB1*04:32 allele.

Internal investigation has confirmed, with the now available rare sample type, that lane 9 is false negative for the DRB1 *04:32 allele in the listed affected kit according to the labeling provided with the kit. Product does not meet specification. However, the primer mix in lane 9 is functioning as originally intended. The DRB1*04:32 allele has a mismatch in the primer's targeted sequence and sample specific functionality could not be confirmed through testing.