Health Products

Enteralite and Infinity Orange Pumps

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer
Manufacturer
Zevx Inc., also trading as Moog Medical Devices Group
A. Enteralite and Infinity Orange PumpsINFKIT2

Introduction

Recalled Products

A. Enteralite and Infinity Orange Pumps

Reason

The recall concerns an error ER13 for Zevex's Moog Enteralite Feeding Pumps manufactured between the dates August 16, 2010 and April 15, 2011.  The error is caused by defective diodes.  Even if the diodes pass the manufacturing controls testing, there still is a possibility of pump failure.  The recall is being initiated to ensure all pumps in field receive the correction.