Health Products

Mobicath Bi-Directional Guiding Sheath

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
W2172712, W2190464, W2196349, W2196351, W2211853
Manufacturer
Greatbatch Medical
A. Mobicath Bi-Directional Guiding SheathD-1400-10, D-1400-11

Introduction

Recalled Products

A. Mobicath Bi-Directional Guiding Sheath

Reason

The manufacturer has observed some anomalies of the inner lumen of these products with the implementation of a new inspection tool during visual inspection. The anomalies may include loose, string-like liner material, liner abrasion (fraying), and/or attached string-like liner material (from scratches or damage on the inner wall of sheath lumen).