Intellivue Products Nov 9, 2012 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer Manufacturer Philips Medical Systems B. Intellivue - Information Center M3145, M3150, M3155 More than 10 numbers, contact manufacturer Manufacturer Philips Medical Systems C. Intellivue - Information Center IX-Software M3290B 5226A05453, 5226A05454, 5226A05514, 5226A08627, 5232A10872, 5232A10873, 5232A10874, 5232A10875, 5232A10876, 5232A10877 Manufacturer Philips Medical Systems A. Intellivue - Information Center Client M3151B
Recalled Products
Intellivue - Information Center Client
Intellivue - Information Center
Intellivue - Information Center IX-Software
Philips healthcare has become aware that an internal audio cable connector may have diminished retention force due to damaged contacts. This diminished retention force may lead to intermittent or loss of audible alarms.