Health Products

Advia Centaur Systems

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
84160044, 84249044, 84631044, 84871044, 86097045, 86415045, 86464045, 87448046, 87929046
Manufacturer
Siemens Healthcare Diagnostics Inc.
B. Advia Centaur XP System - HBC Total Assay07566733
84160044, 84249044, 84631044, 84871044, 86097045, 86415045, 86464045, 87448046, 87929046
Manufacturer
Siemens Healthcare Diagnostics Inc.
C. Advia Centaur CP System - HBC Total Assay07566733
84160044, 84249044, 84631044, 84871044, 86097045, 86415045, 86464045, 87448046, 87929046
Manufacturer
Siemens Healthcare Diagnostics Inc.
A. Advia Centaur System - HBC Total Assay07566733

Introduction

Recalled Products
  1. Advia Centaur System - HBc Total Assay
  2. Advia Centaur XP System - HBc Total Assay
  3. Advia Centaur CP System - HBc Total Assay

Reason

Siemens has received complaints regarding a perceived increased rate of reactive results with specific Advair Centaur HBCT Assay kit lots, that do not correspond with other HBV Serological Markers. HBCT Readypack lot numbers ending in 044, 045, and 046 do not meet the specificity claims in the IFU. Product code 07566733 is used with the Advia Centaur, Advia Centaur XP, and Advia Centaur CP systems.