· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Product Safety
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
080310 to 310812D. Aquavent Heated Breathing Circuits
Manufacturer
Armstrong Medical Ltd.
AMVC1774/101, AMVC1774/102, AMVC1774/105, AMVC1774/106, AMVC1775/101, AMVC1775/102, AMVC1775/106, AMVC1775/108
080310 to 310812E. Aquavent Bipap Limb for Non-Invasive Ventilation
Manufacturer
Armstrong Medical Ltd.
AMVC1792/130
080310 to 310812B. Aquavent CPAP Limb, Adult, Heated
Manufacturer
Armstrong Medical Ltd.
AMCP1409/003, AMCPUK01109
080310 to 310812C. Aquavent Limb for Oxygen Therapy, Adult, Heated
Manufacturer
Armstrong Medical Ltd.
AMCP1409/004
080310 to 310812A. CPAP System - Circuits
Manufacturer
Armstrong Medical Ltd.
AMCPSI01052, AMCPUK01103

Introduction

Recalled Products
  1. CPAP System - Circuits
  2. Aquavent CPAP Limb, Adult, Heated
  3. Aquavent Limb for Oxygen Therapy, Adult, Heated
  4. Aquavent Heated Breathing Circuits
  5. Aquavent Bipap Limb for Non-Invasive Ventilation

Reason

Armstrong Medical Ltd. is investigating the heater wire retainer (located in the heated limbs of adult corrugated heated ventilator and CPAP breathing circuits as a means to locate the heater wire in a fixed position).

The retainer is white or off-white in color and made from polymeric (plastics) material. If breakage of this component were to occur there could be a possible inhalation risk to the patient.