Health Products

Unicel DXH Slidemaker Stainer Coulter Cellular Analysis system, Software Version 1.5.1.0

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Beckman Coulter Inc.
A. Unicel DXH Slidemaker Stainer Coulter Cellular Analysis system, Software Version 1.5.1.0775222

Introduction

Recalled Products

A. Unicel DXH Slidemaker Stainer Coulter Cellular Analysis system, Software Version 1.5.1.0

Reason

Beckman Coulter Inc. (BEC) has confirmed that the system manager on the Unicel DXH Slidemaker Stainer Coulter Cellular Analysis system, software version 1.5.1.0, will not allow the customers to change the time and date backward when following the Instruction for Use (IFU). This affects the ability of an operator to make the necessary time change adjustments for the changes from Daylight Savings Time (DST) to Standard Time.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime