AirLife Isothermal Breathing Circuit Products Jun 20, 2012 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public (GP), Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Brand name Companies Model or catalog number All less than Lot #000415257 or D. AirLife Isothermal Breathing Circuit Infant Respiratory Circuit MR850 Manufacturer carefusion RT4851-00, RT4851-12, RT4851-18 All less than Lot #000415257 or B. AirLife Isothermal Breathing Circuit Infant Respiratory Circuit (Heated or Non-Heated) Manufacturer carefusion More than 10 Numbers, contact manufacturer All less than Lot #000415257 or C. AirLife Isothermal Breathing Circuit Infant Respiratory Circuit Heated Manufacturer carefusion 1815-409 All less than Lot #000415257 or A. AirLife Isothermal Breathing Circuit Accessory- Infant Patient Connector Manufacturer carefusion 5925-504
Recalled Products
AirLife Isothermal Breathing Circuit Accessory- Infant Patient Connector
AirLife Isothermal Breathing Circuit Infant Respiratory Circuit (Heated or Non-Heated)
AirLife Isothermal Breathing Circuit Infant Respiratory Circuit Heated
AirLife Isothermal Breathing Circuit Infant Respiratory Circuit MR850
The y adapter within the circuit may spontaneously crack causing a leak in the closed ventilation system, the leak may cause a loss of tidal volume being delivered to the patient and the ventilator alarm may sound. The leak could lead to a decrease in the tidal volume delivered to the ventilated patient.