Health Products

Access HLH Calibrators (Access System)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
008546, 011895, 014099, 022519, 109784, 113689, 118299, 122157, 220438,B. Access HLH Calibrators (Unicel DXI 600)
Manufacturer
Beckman Coulter Inc.
33515
008546, 011895, 014099, 022519, 109784, 113689, 118299, 122157, 220438,C. Access HLH Calibrators (Unicel DXI 800)
Manufacturer
Beckman Coulter Inc.
33515
008546, 011895, 014099, 022519, 109784, 113689, 118299, 122157, 220438, 916872A. Access HLH Calibrators (Access System)
Manufacturer
Beckman Coulter Inc.
33515

Introduction

Recalled Products
  1. Access HLH Calibrators (Access System)
  2. Access HLH Calibrators (Unicel DXI 600)
  3. Access HLH Calibrators (Unicel DXI 800)

Reason

The affected lots do not meet the stated expiration date claim. As a result, the quality control and patient sample results generated while using the affected lot numbers might be falsely elevated by as much as 13%.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime