Health Products

Base for Terumo Advanced Perfusion System 1

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Terumo Cardiovascular Systems Corp.
A. Base for Terumo Advanced Perfusion System 1801763

Introduction

Recalled Products

A. Base for Terumo Advanced Perfusion System 1

Reason

Terumo Cardiovascular Systems (Terumo CVS) has become aware of a remote possibility that Terumo Advanced Perfusion System 1 could experience a spontaneous loss of power. Terumo CVS is issuing this advisory because the user's manual for the Terumo Advanced Perfusion System 1 does not provide user instruction for responding to such an event.