A. BD Vacutainer Plus 2.7 mL Citrate Blood Collection Tube
Manufacturer
Becton Dickinson and Company
363083
Introduction
Recalled Products
A. BD Vacutainer Plus 2.7 mL Citrate Blood Collection Tube
Reason
There is evidence that in a portion of the affected lot there is a cross-contamination with lithium heparin, in a concentration not exceeding 1.2 iu per milliliter in plasma obtained from properly filled tubes (note: this value was obtained from healthy subjects using a limited number of tubes from the affected lot). The internal investigation completed by bd indicates that this contaminant can adversely affect coagulation results. Erroneous results may lead to inappropriate decisions regarding patient care. For a full listing of assays that may be adversely affected by the lithium heparin contaminant, please contact the manufacturer.