Life Pulse Patient Circuits Dec 12, 2012 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type I
Lot or serial number Companies Brand name Model or catalog number More than 10 numbers, contact manufacturer Manufacturer Bunnell Inc. A. Life Pulse Patient Circuits 937
Recalled Products
Life Pulse Patient Circuits
A specific lot of heater wire insulation may fail by melting causing sparking and smoking. The resulting circuit failure could result in patient injury or death.