Health Products

MobileDiagnost WDR

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
N/AA. MobileDiagnost WDR
Manufacturer
Philips Medical Systems DMC GMBH
9890 010 8952X

Introduction

Recalled Products
  1. MobileDiagnost WDR

Reason

  1. When the operator moves an image to another view, the electronic marker might be placed at the wrong anatomical position in the resulting image.
  2. If by an electrical connection failure between motor speed sensor and digital motion control board or a related software defect, the movement speed regulation does not work as intended, the device may move with increasing speed and/or into a wrong direction after releasing the dead man switch (handle bar) during 1.5 seconds maximum until the brake is activated.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime