Health Products

SOMATOM Products

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Any Serial Number running Software Version VA20, VA27, VA34 or VA40
Manufacturer
Siemens Ag
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. SOMATOM Definition – AS – Main Unit8098027
Any Serial Number running Software Version VA20, VA27, VA34 or VA40
Manufacturer
Siemens Ag
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C. SOMATOM Definition Flash Sys – Main Unit10430603
Any Serial Number running Software Version VA20, VA27, VA34 or VA40
Manufacturer
Siemens Ag
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A. SOMATOM Definition – Main Unit77 40 769

Introduction

Recalled Products

A. SOMATOM Definition – Main Unit

B. SOMATOM Definition – AS – Main Unit

C. SOMATOM Definition Flash Sys – Main Unit

Reason

There is a highly sporadic malfunction that concerns the 3D reconstruction functionality of the SOMATOM Systems. This problem may occur during the process of a 3D reconstruction job planning and editing within a specific funtionality, while another functionality ("preview image") is activated. The problem presents itself as an error message window and the system needs to be rebooted.