Health Products

Dimension Products – Total Iron Binding Capacity (IBCT)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public (GP), Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 Numbers, Contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
B. Dimension Xpand System - IBCTDF84
More than 10 Numbers, Contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
C. Dimension EXL System – IBCTDF84
More than 10 Numbers, Contact manufacturer
Manufacturer
Siemens Healthcare Diagnostics Inc.
A. Dimension Clinical Chemistry – IBCTDF84

Introduction

Recalled Products

A. Dimension Clinical Chemistry – IBCT
B. Dimension Xpand System – IBCT
C. Dimension EXL System – IBCT

Reason

Through monitoring of method Performance Siemens has determined that the analytical sensitivity claim of 6 mcg/dl (1.1 mcmol/l) stated in the instructions for use is no longer valid for Dimension Total Iron Binding Capacity (ibct), and is modifying that claim to 36 mcg/dl (6.44 mcmol/l).