Health Products

Dimension Products – Iron Calibrations

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2DN060, 2FN063, 2GN068, 2JN074
Manufacturer
Siemens Healthcare Diagnostics Inc.
B. Dimension Xpand System – Iron AssayDC85
2DN060, 2FN063, 2GN068, 2JN074
Manufacturer
Siemens Healthcare Diagnostics Inc.
C. Dimension EXL System - Iron AssayDC85
2DN060, 2FN063, 2GN068, 2JN074
Manufacturer
Siemens Healthcare Diagnostics Inc.
A. Dimension Clinical Chemistry System – (Iron Flex Reagent Cartridge)DC85

Introduction

Recalled Products
  1. Dimension Clinical Chemistry System – (Iron Flex Reagent Cartridge)
  2. Dimension Xpand System – Iron Assay
  3. Dimension EXL System – Iron Assay

Reason

Complaints received regarding iron calibrations outside of acceptable Calibration Slope (M) values (0.97-1.03) when using Iron Calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints received.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime