SenTec Digital Monitoring System (SDMS)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1D58, 1D80
Manufacturer
Sentec AG
A. SenTec Digital Monitoring System (SDMS)VS-MCI

Introduction

Recalled Products
  1. SenTec Digital Monitoring System (SDMS)

Reason

Due to a manufacturing defect that occurred for certain inserts, the new electrolyte cannot be expelled due to the fact that the hole for expelling the electrolyte is closed. If a membrane changer with the defective insert is used to change the membrane, as a consequence the sensor will be remembraned without electrolyte. The PC02 Sensor performance will be deteriorated without the presence of electrolyte, giving unstable or questionable results.