Access Immunoassay Systems (by Beckman Coulter)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
120380, 122262
Manufacturer
Beckman Coulter Inc.
B. Access Immunoassay Sys - PAPP-A (Pregnancy Associated Plasma Protein) Assay120380, 122262
120380, 122262
Manufacturer
Beckman Coulter Inc.
C. Unicel DXI 600 Access Immunoassay System - PAPP-A SystemA48573
120380, 122262
Manufacturer
Beckman Coulter Inc.
A. Unicel DXI 800 Access Immunoassay System - PAPP-A AssayA48573

Introduction

Recalled Products
  1. Unicel DXI 800 Access Immunoassay System - PAPP-A Assay
  2. Access Immunoassay Sys - PAPP-A (Pregnancy Associated Plasma Protein) Assay
  3. Unicel DXI 600 Access Immunoassay System - PAPP-A System

Reason

Beckman Coulter, Inc. (BEC) has identified that two access PAPP-A Quality Control (QC) lots (120380, 122262) may generate mean values below the mean values stated on the QC cards. The average difference observed in BEC's internal investigation was 11% below the stated mean. The results observed may vary based on the combination of reagent and calibrator lots in use. Reduced QC could lead to repeat testing and a potential delay in reporting patient results.