Health Products

Full Auto Ref-Keratometer

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
200004
Manufacturer
Canon Inc.
A. Full Auto Ref-KeratometerRK-F2

Introduction

Recalled Products
  1. Full Auto Ref-Keratometer

Reason

Potential variations in values measured in the Ref mode (Refractive Power) were reported with respect to the firmware included in RK-F2 Full Auto Ref-Keratometer. Canon Inc. has determined that there are potential variations in the optical system switching time between the Alignment Motor and Measurement Period and, therefore, fluctuations in measurement values may occur. Canon Inc. will resolve this potential issue by upgrading the firmware in the RK-F2 Auto Ref-Keratometer.