Health Products

Animas 2020 Insulin Infusion Pump

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 100 numbers, contact manufacturer.A. Animas 2020 Insulin Infusion Pump
Manufacturer
Animas Corporation
100-380-51

Introduction

Recalled Products
  1. Animas 2020 Insulin Infusion Pump

Reason

A small number of Animas 2020 Insulin Infusion Pumps are potentially affected by a misalignment of a component on the printed circuit board assembly. This component misalignment results in the Insulin Pump potentially measuring force conditions incorrectly, resulting in false warnings and/or alarms. Should the patient complete a load sequence or prime the insulin infusion pump and does not disconnect from the infusion set from the body, the insulin infusion pump may dispense all of the insulin in the cartridge into the patient. There is no safety risk as long as the patient follows the instructions for use and always disconnects the infusion set prior to priming the pump, changing the cartridge, and rewinding the motor.