Health Products

Unicel Clinical Analyzers and Immunoassay Systems

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
900359
Manufacturer
Beckman Coulter Inc.
H. Unicel DXI 600 Access Immunoassay System - Instrument Class IIA71461
900359
Manufacturer
Beckman Coulter Inc.
I. Unicel DXI 800 Access Immunoassay - Analyzer - Class 3A71461
900227
Manufacturer
Beckman Coulter Inc.
F. Unicel DXI 600 Access Immunoassay System4771
900359
Manufacturer
Beckman Coulter Inc.
G. Unicel DXI 600 Access Immunoassay System - Instrument Class 34773
More than 10 numbers, contact manufacturer.D. Unicel DXI 600 Access Immunoassay System - Instrument with Dual Gantry Class 2973100
More than 10 numbers, contact manufacturer.
Manufacturer
Beckman Coulter Inc.
E. Unicel DXI 600 Access Immunoassay System - Instrument with Dual Gantry Class 3973100
900227
Manufacturer
Beckman Coulter Inc.
B. Unicel DXC 860i Synchron Access Clinical AnalyzerA30260
More than 10 numbers, contact manufacturer.
Manufacturer
Beckman Coulter Inc.
K. Unicel DXI 800 Access - Analyzer - Class II973100
600785
Manufacturer
Beckman Coulter Inc.
C. Unicel DXC 880i Synchron Access Clinical Analyzer4768
900227
Manufacturer
Beckman Coulter Inc.
A. Unicel DXC 660i Synchron Access Clinical AnalyzerA30260
900227
Manufacturer
Beckman Coulter Inc.
J. Unicel DXI 800 Access Immunoassay SystemA30260

Introduction

Recalled Products
  1. Unicel DXC 660i Synchron Access Clinical Analyzer
  2. Unicel DXC 860i Synchron Access Clinical Analyzer
  3. Unicel DXC 880i Synchron Access Clinical Analyzer
  4. Unicel DXI 600 Access Immunoassay System – Instrument with Dual Gantry Class 2
  5. Unicel DXI 600 Access Immunoassay System – Instrument with Dual Gantry Class 3
  6. Unicel DXI 600 Access Immunoassay System
  7. Unicel DXI 600 Access Immunoassay System – Instrument Class 3
  8. Unicel DXI 600 Access Immunoassay System – Instrument Class II
  9. Unicel DXI 800 Access Immunoassay - Analyzer – Class 3
  10. Unicel DXI 800 Access Immunoassay System
  11. Unicel DXI 800 Access - Analyzer – Class II

Reason

As a part of the 10,000-Test Maintenance Procedure, which occurs on average once a month in the DXI family of instruments, customers are required to replace aspirate and dispense probes and tubing. The dispense and aspirate tubing can become pinched or damaged as a result of incorrect customer installation of the tubing after probe maintenance or replacement, which may prevent the DXI and DXC Systems from washing reaction vessels properly and may lead to erroneous test results.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime