Health Products

Cutting and Buttings Burrs and Micromax System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Product Safety, Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All Lots
Manufacturer
The Anspach Effort Inc.
B. Buttings Burrs - Fluted BallMCA5-15SB
All Lots
Manufacturer
The Anspach Effort Inc.
C. Micromax System - Micro Curved AttachmentMCA
All Lots
Manufacturer
The Anspach Effort Inc.
A. Cutting Burrs - Diamond BullMore than 10 numbers, contact manufacturer.

Introduction

Recalled Products
  1. Cutting Burrs - Diamond Bull
  2. Buttings Burrs - Fluted Ball
  3. Micromax System - Micro Curved Attachment

Reason

The current directions for use for the MCA Cutting Burrs and MCA Micro Curved Attachment do not provide clear details regarding the intended use specifically for the MCA burrs. The use of the device in dense bone removal may result in cutter fracture.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime