Health Products

Instrument Manager Software (Using Roche Cobas E411 Driver)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Version 8.00.0001
Manufacturer
Data Innovations LLC
A. Instrument Manager Software (Using Roche Cobas E411 Driver)8.00.0001

Introduction

Recalled Products

A. Instrument Manager Software (Using Roche Cobas E411 Driver)

Reason

When using the Roche Cobas E411 (RCHCE41I) driver version 8.00.0001 or 8.00.0002, if there is a barcode read error, the current sample may be identified as the last sample ID that was in that rack or disk position potentially causing sample mismatch. If the last sample had any pending tests in the orders database, test selection would be downloaded to the instrument with the incorrect sample ID, and results would be reported on the wrong sample.