Lot or serial numberBrand nameCompaniesModel or catalog number
255406B. Advia 1200 Chemistry System – Alkaline Phosphatase (ALPAMP) Assay 
Manufacturer
Siemens Healthcare Diagnostics Inc.
03035814
255406A. Advia Chemistry System – Alkaline Phosphatase (ALPAMP) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
03035814

Introduction

Recalled Products
  1. Advia Chemistry System – Alkaline Phosphatase (ALPAMP) Assay
  2. Advia 1200 Chemistry System – Alkaline Phosphatase (ALPAMP) Assay

Reason

There is an increased incidence of absorbance flags (u, u, or ////) on some ALPAMP and ALPDEA reagent wedges, which in some cases can impede reporting of results. No QC shift or discrepant patient results are observed.