Health Products

Bluephase

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
603752BU: 400000 to 400705
Manufacturer
Ivoclar Vivadent AG
A. Bluephase607920BU

Introduction

Recalled Products

A. Bluephase

Reason

In two cases to date, a very rare combination of swelling of the battery cell with a ridge on the inside plastic housing of the battery caused a chain reaction, allowing energy in the handpiece battery to be released, thus destroying the device and causing smoke damage.
Get told about the next health product recall.
Get health product recall alerts