The Alaris Pump Module Model 8100 manufactured between August 2010 and July 2011 may experience an intermittent motor stall during infusion. The investigation revealed that a mechanical binding within the mechanism assembly is resulting in a motor stall. When a motor stall occurs, the Alaris PC Unit and the Alaris Pump Module Display the Visual Error Code 242.4030 with an audible alarm and is followed by a termination of infusion.