The Monthly Cleaning Procedure (MCP) maintenance, under certain circumstances, may result in residual cleaning solution (Bleach) in the Reagent Probe 1 (RP1) fluidics and the water reservoir causing false results. Tests affected are Total Human Chorionic Gonadotropin (THCG), Progesterone and Prostate Specific Antigen (PSA). The Daily Cleaning Procedure (DCP) under certain circumstances may result in residual cleaning solution (Bleach) in the water line causing false results that use water for test processing. Test affected is Toxoplasma M (Toxo M).