Dimension Hemoglobin A1C Flex Reagent

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
GA3113, GA3134, GA3141, GA3169, GA3099, GA3162
Manufacturer
Siemens Healthcare Diagnostics Inc.
A. Dimension Hemoglobin A1C Flex ReagentDF105A

Introduction

Recalled Products

A. Dimension Hemoglobin A1C Flex Reagent

Reason

There has been a change to the acceptable range for the 'E' calibration coefficient guideline. The new guideline range is 5 to 8 (conventional units). The guideline range for Lots prior to GA3099 is 6 to 9. The E coefficient range is a criterion used by customers to assess the validity of their calibration. Siemens has also noted that some users are not updating Scaler Values when setting up a calibration (this was implemented beginning with Lot GA3009), resulting in inaccurate values for both patients and controls.