There has been a change to the acceptable range for the 'E' calibration coefficient guideline. The new guideline range is 5 to 8 (conventional units). The guideline range for Lots prior to GA3099 is 6 to 9. The E coefficient range is a criterion used by customers to assess the validity of their calibration. Siemens has also noted that some users are not updating Scaler Values when setting up a calibration (this was implemented beginning with Lot GA3009), resulting in inaccurate values for both patients and controls.