Health Products

DRX-Evolution System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
4236
Manufacturer
Carestream Inc.
D. DRX-Evolution / with Motor, Fixed Wall8009839
4031, 4076
Manufacturer
Carestream Inc.
B. DRX-Evolution / Premium OTC FF Table1268457
4670, 4199, 4569, 4588, 4570
Manufacturer
Carestream Inc.
C. DRX-Evolution / Motorized OTC FF Wall1731678, 8001174, 8716938
4292, 4208
Manufacturer
Carestream Inc.
A. DRX-Evolution / Premium OTC FF Wallstand1001700, 1001981

Introduction

Recalled Products
  • A. DRX-Evolution / Premium OTC FF Wallstand
  • B. DRX-Evolution / Premium OTC FF Table
  • C. DRX-Evolution / Motorized OTC FF Wall
  • D. DRX-Evolution / with Motor, Fixed Wall

Reason

Carestream has identified that the DRX-Evolution system may exhibit a gradual downward movement of the Overhead Tube Crane (OTC) in the event of a power loss or power outage.