Health Products

Ameriwater MRO Dialysis RO+ System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
31000-31079
Manufacturer
Ameriwater
A. Ameriwater MRO Dialysis RO+ System00MRO120, 00MRO121, 00MROS20, 00MROS21, MRO1, MROS

Introduction

Recalled Products

A. Ameriwater MRO Dialysis RO+ System

Reason

Ameriwater has identified a potential problem with the Ameriwater Portable RO+ System Model MROS20, MROS21, MRO120 and MRO121 with serial numbers between 31000 and 31079. Ameriwater specified 250 Volt 15 Amp fuses, but the supplier of the motor starter mistakenly installed 32 Volt fuses in the assembly.