Health Products

Dash Patient Monitor

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N/A
Manufacturer
GE Medical Systems Information Technologies, Inc.
B. Dash 5000 Patient MonitorDASH5-XXXX-XXXX-XXXX
N/A
Manufacturer
GE Medical Systems Information Technologies, Inc.
C. Dash 4000 Patient MonitorDASH4-XXXX-XXXX-XXXX
N/A
Manufacturer
GE Medical Systems Information Technologies, Inc.
A. Dash 3000 Patient MonitorDASH3-XXXX-XXXX-XXXX

Introduction

Recalled Products
  1. Dash 3000 Patient Monitor
  2. Dash 5000 Patient Monitor
  3. Dash 4000 Patient Monitor

Reason

The Dash monitor may exhibit an unexpected loss of Custom Automatic NBP Measurement Cycling. When the Custom Automatic NBP Measurement Cycling turns off, the auto message is no longer displayed in the NBP Parameter Window at the patient bedside monitor. The timestamp of the last NBP will remain displayed and not reset, if no NBP Monitoring has taken place after 2 hours in Adult ICU Mode, 15 minutes in the Operating Room Mode or 12 hours in Neonatal-ICU Mode.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime