Health Products

Leica RTU-TDT-339, 7mL

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
6001845, 6009645, 804804, 804805, 804806, 804809, 804810, 804812
Manufacturer
Leica Biosystems Newcastle Ltd.
A. Leica RTU-TDT-339, 7mLRTU-TDT-339

Introduction

Recalled Products
  1. Leica RTU-TDT-339, 7mL

Reason

The recall was intiated as a result of findings that the product may not retain stability through out it's entire 36 month shelf life. The investigation has shown that there is a link between the age of product and the staining intensity and it was found the labelled 36 month shelf life was not appropriate.