Health Products

Immulite Systems

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Lot 315 to 335
Manufacturer
Siemens Healthcare Diagnostics Products Limited
B. Immulite System IGF-1 (Insulin-like Growth Hormone) AssayLKGF1
Lot 315 to 335
Manufacturer
Siemens Healthcare Diagnostics Products Limited
C. Immulite 1000 System IGF-1 AssayLKGF1
Lot 441 to 492
Manufacturer
Siemens Healthcare Diagnostics Products Limited
A. Immulite 2000 System IGF-1 (Insulin-like Growth Hormone) AssayL2KGF2

Introduction

Recalled Products
  1. Immulite 2000 System IGF-1 (Insulin-like Growth Hormone) Assay
  2. Immulite System IGF-1 (Insulin-like Growth Hormone) Assay
  3. Immulite 1000 System IGF-1 Assay

Reason

There is a positive shift in patient medians of approximately 20% with the igf-I assay. An initial positive shift of approximately 15% occurred with immulite 2000/immulite 2000 xpi kit lot 441 (released december 2009), and in mid-2011 an additional positive shift of approximately 5% was observed with immulite 2000/immulite 2000 xpi kit lot 469 (released july 2011), leading to the total positive shift of approximately 20%.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime