Health Products

Amvisc and Amvisc Plus

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 Lot numbers, Contact Manufacturer
Manufacturer
Bausch & Lomb Inc.
B. Amvisc Plus60051, 60081
More than 10 Lot numbers, Contact Manufacturer
Manufacturer
Bausch & Lomb Inc.
A. Amvisc59081L

Introduction

Recalled Products
  1. Amvisc
  2. Amvisc Plus

Reason

Some disposable cannulas provided with the Amvisc or Amvisc Plus OVD are not properly engaging to the Luer-lock on the Amvisc and Amvisc Plus sterile glass syringes. They may leak viscoelastic material or detach from the syringe during injection.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime