52803UN12, 62299UN12, 85773UN12A. Aeroset System – Phenobarbital Reagent
Manufacturer
Abbott Laboratories Diagnostic Division
1E08-01

Introduction

Recalled Products
  1. Aeroset System – Phenobarbital Reagent

Reason

Abbott Diagnostics is recalling clinical chemistry phenobarbital reagent lots 62299un12 and 85773un12 due to imprecision issues which increase with the age of the reagent lot. The imprecision of the phenobarbital assay is less than or equal to 7% total cv%, however, internal testing confirmed cvs up to approximately 12% for in-date reagent lots over five months old. Although lot 52803un12 is now expired, it exhibited similar imprecision issues. As a precaution, future replacement lots will have reduced dating until root cause is identified and preventive measures are implemented.