Health Products

Instatrak 3500 Plus System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N/A
Manufacturer
GE OEC Medical Systems Inc.
A. Instatrak 3500 Plus System1006236-001

Introduction

Recalled Products
  1. Instatrak 3500 Plus System

Reason

The reason the recall was intiated was because several potential safety issues have been identified with the use of the device. The navigation system may lock up or restart unexpectedly. If two scans from the same modality and same patient are transferred to the system, they may combine into one scan, and the transmitter arm movement after registration may result in inaccuracy and users may experience difficulty calibrating the device.