Health Products

Multidiagnost Eleva-Image Intensifier and Multidiagnost Eleva-Flat Detector

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
N/AB. Multidiagnost Eleva-Flat Detector – Image Detector
Manufacturer
Philips Medical Systems Nederland B.V.
0708MD02
N/AA. Multidiagnost Eleva-Image Intensifier – TV Image Detector
Manufacturer
Philips Medical Systems Nederland B.V.
0708MD02

Introduction

Recalled Products
  1. Multidiagnost Eleva-Image Intensifier – TV Image Detector
  2. Multidiagnost Eleva-Flat Detector – Image Detector

Reason

Due to an incorrect algorithm, the Air Kerma (AK) and Air Kerma Rate Values are not displayed correctly. The displayed AK is using an incorrect point of reference. This is a non-compliance per the Radiation Emitting Devices Act (R.S.C., 1985, section 6).