Health Products

Urea Nitrogen (for the AU 480 / AU 680 Analyzers Only)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All Lots
Manufacturer
Beckman Coulter, Inc.
A. Urea Nitrogen (for the AU 480 / AU 680 Analyzers Only)OSR6134

Introduction

Recalled Products
  1. Urea Nitrogen (for the AU 480 / AU 680 Analyzers Only)

Reason

Some customers may see an increase in calibration factor for urea nitrogen reagent osr6134 and osr6234 on the au480/au680 analyzers, which in turn may lead to calibration failures that could delay quality control or patient sample analysis.