One complaint has been received. DRB1 High Res SSP Unitray Kits have a false negative in lane 44 for DRB1*04:33 allele.
Internal investigation confirmed by sequence based typing that the customer had a DRB1*04:33 allele. Therefore lane 44 has a confirmed false negative in the affected kits according to the labeling. However, the primer mix in lane 44 is functioning as originally designed. The DRB1*04:33 allele has a mismatch in the primer's targeted sequence and sample specific functionality could not be confirmed through testing.