Health Products

Lumex Patriot Homecare Bed

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 Numbers, Contact manufacturer
Manufacturer
Basic American Div. Graham Field
A. Lumex Patriot Homecare Bed690-7004-000

Introduction

Recalled Products
  1. Lumex Patriot Homecare Bed

Reason

Correction and removal is being conducted for the Lumex Patriot Full-Electric Homecare "bed ends". There is potential to provide room for the leg pin to move during use. If failure were to occur, the bed could descend rapidly causing non-typical patient movement and could pose a risk of injury to the patient and others.